Ceyear 4024E 유지 관리 매뉴얼 - 페이지 14

{카테고리_이름} Ceyear 4024E에 대한 유지 관리 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Ceyear 4024E 19 페이지. Spectrum analyzer

Ceyear 4024E 유지 관리 매뉴얼
cable with the PC through the GPIB card, and connect 4024 to be tested with the PC through the
direct-connected network cable. It is as shown in the figure:
b)
Open the temperature compensation software TempComp.exe, and the software interface is as
follows:
Click the Set Address button, and the communication address setting dialog will pop up. As shown in the
following figure, the GPIB address of the signal source to be connected, the GPIB Address of the power
meter and the IP Address of the spectrum analyzer can be set. The address of the signal source and
power meter can be inquired through 【System】 menu. If the GPIB number of the signal source is 19,
no modification is required; similarly, if the GPIB number of the power meter is 13, no modification is
required. The IP address of the spectrometer can be inquired through 【System】->[Network Config].
GPIB card
Signal source
2.4 mm cable
PC
Fig. 9 Schematic Diagram of Connection
Fig. 10 TempComp Interface
GPIB cable
Adapter
Power meter
Power
Adapter
sensor