Calibrating your Orion Nano
The Orion Nano is a precision instrument, and regular calibration is recommended
with frequent use. Here's how to perform a smart calibration for your Orion Nano.
Press the
Dose
rapidly 5 times in the
weighing mode to
enter the user
calibration mode.
Note:
You may use a 100 g or 500 g weight for calibration. To ensure accuracy, use Acaia
calibration weights or F1 or F2 Class calibration weights.
100G
While LED display shows
[ CAL ] and then [ 100 ].
Place a 100 g calibration
weight and wait for the
scale to execute the
calibration procedure.
The Display shows
[ -End- ] when the
calibration procedure
is completed.