AW-Lake EDG1 설치 운영 및 유지보수 매뉴얼 - 페이지 9
{카테고리_이름} AW-Lake EDG1에 대한 설치 운영 및 유지보수 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. AW-Lake EDG1 18 페이지. The edge flow sensor with bluetooth
Output Calibration
Warning modifying these outputs could result in inaccurate analog signals.
This menu is provided to adjust the EDGE analog output to correspond correctly with the
analog output of the input device.
* This menu is intended for use with slight output differences between the EDGE output and customer
input device. These values are factory calibrated before shipping to match meters that are calibrated
yearly.
Select Calibration Type: From this menu both mA and Voltage
outputs may be calibrated. (Figure 6)
Channel A – Voltage output
Channel B – mA output
1. Select which output type to be calibrated; mA or Voltage
2. Press Set Output Low, then enter in the value on user's
device into Record Reading
a. Ex. If EDGE is being calibrated to a multimeter,
enter in reading from multimeter to Record
Reading
3. Press Record button
4. Now press Set Output High to calibrate the high end of the output
5. Enter in High reading from input device to Record Reading
6. Press Record button
7. Verify both Calibration high and low values are correct and press Set Calibration
8. Use back button on phone to navigate to the main menu
*Warning do not do 2 calibrations in a row. After calibration is finished, exit the app and restart
the device.
Figure 6:
Output Calibration
9 |
P a g e