CHS MED-RX Series 사용 지침

{카테고리_이름} CHS MED-RX Series에 대한 사용 지침을 온라인으로 검색하거나 PDF를 다운로드하세요. CHS MED-RX Series 2 페이지. Instruments

 
CLEANING AND MAINTENANCE 
A) Rinsing 
Directly  after  surgery,  rinse  instruments  under  hot
running water, removing all body fluids and debris. 
B) Disinfecting  (the  protection  for  medical  personnel 
from accidental contamination during cleaning) 
To  avoid  blood  and  other  proteins  from  sticking  to 
instrument surfaces, an enzymatic cleaner bath (soaking) 
should  be  used  on  all  instruments.  After  soaking  for  a 
minimum of 10 minutes, rinse all instruments in running 
tap water. 
Immerse  instruments  completely  in  hospital  approved 
disinfectant for an additional 10 minutes then rinse again 
under running water. 
DO  NOT  USE  BLEACH.  Exposure  to  bleach  will  damage 
instruments. 
C) Cleaning 
Instruments should be submerged in a solution of water 
and neutral pH (7) detergent.
Ultrasonic Cleaning  
Place instruments in an open position into the ultrasonic 
cleaner.  Make  sure  all  "sharp"  (i.e.  scissors,  knives, 
blades,  etc.)  do  not  touch  other  instruments.  All 
instruments 
must 
be 
completely 
Instruments  should  be  processed  in  the  ultrasonic 
cleaner  for  the  full  recommended  cycle  time,  which  is 
typically 5‐10 minutes. A lid should cover the ultrasonic 
cleaner  during  the  operation  to  avoid  splashing.  Never 
place  different  metals  in  the  same  cleaning  cycle.  For 
example, stainless steel and copper, etc.  
Change 
solution 
frequently 
per 
recommendations.  Rinse  all  instruments  after  an 
ultrasonic  cleaning  with  water  to  remove  the  excess 
cleaning solution. 
Automatic Washing Sterilization 
Follow all manufacturers' recommendations.  Ensure all 
instruments  are  lubricated  after  and  before  the 
sterilization cycle. 
Manual Cleaning 
Ultrasonic cleaning is the preferred method of cleaning 
and strongly recommended to ensure all dirt, debris and 
are  fully  removed  –  particularly  with  instruments  with 
hinges,  locks,  and  other  moving  parts.    If  ultrasonic 
cleaning is not available, follow the below instructions. 
a) Use  stiff  nylon  cleaning  brushes.    Do  not  use  steel
wool  or  wire  brushes  unless  specifically  recommended 
for cleaning brushes for instrument‐serrated areas or on 
bone files, burs or on stained areas in knurled handles.
Recommended  Cleaning  Procedure  for  Cleaning 
Maintenance and Sterilization of Surgical Instruments 
b) Use only neutral pH (7) detergents, low pH (acidic less 
than 6 pH) detergents will cause breakdown of stainless 
steel  protective  surface  (pitting)  and  black  staining.
Higher  pH  detergents  (alkaline  –  more  than  8  pH)  will
cause  a  surface  deposit  of  a  brown  stain  (phosphates), 
which will also interfere with a smooth operation of the 
instrument. Most brown stains are not rust, but merely a 
high  pH  surface  (phosphate)  deposit  and  can  easily  be 
removed with a stain remover.
c) Brush delicate instruments carefully.  These should be
handled separately from all other general instruments. 
d) Make  sure  all  instrument  surfaces  are  visibly  clean 
and free from stains and tissue. The stain remover will 
aid in the elimination of residue stains.
Distributed in Canada by:
Instruments – Instructions for Use
 
*French instructions to follow/* Mode d'emploi en français à suivre
Each  instrument  should  be  inspected  for  functionality 
and  general  condition,  this  would  include,  but  is  not 
limited to: 
 Forceps have properly aligned tips.
 Hemostats and needle holders should not show light 
between  the  jaws,  when  closed  in  the  first  ratchet 
position  (hemostats  may  show  a  small  open  space 
halfway in from the closed tips), lock and unlock easily 
at joints and make sure they are not too loose.  Check
needle holders for wear on jaw surfaces.
 Check suction tubes to ensure they are clean inside.
 Ensure biopsy punches are clean by punching a hole 
into tissue paper. 
 Check working functions of all retractors. 
 Make  sure  that  all  cutting  instruments  and  knives
have sharp undamaged blades.
e) After  scrubbing,  rinse  all  instruments  thoroughly 
under  running  water.  While  rinsing,  open  and  close 
scissors,  hemostats,  needle  holders  and  other  hinged 
instruments  to  make  sure  the  hinged  areas  are  rinsed 
out, as well as the outside of the instruments.
D) After Cleaning 
If instruments are to be stored, let them air dry then 
store them in a clean and dry area.
submerged. 
E) Autoclaving 
Lubricate all instruments that have any "metal to metal"
action (i.e. scissors, hemostats, etc.). Non‐silicone, 
water‐soluble surgical lubricants, such as spray lube is
recommended. Do not sure industrial lubricants. Put 
instruments up for sterilization either individually or in 
sets.
manufacturer 
For Individual Instruments  
Disposable paper or plastic pouches are ideal. Make sure 
you  use  a  wide  enough  pouch  (4  inches  or  wider)  for 
instruments  with  ratchet  locks  (i.e.  hemostats  and 
needle holders, etc.) so the instrument can be sterilized 
in  an  open  (unlocked)  position.  Locking  instruments 
during autoclaving will result in cracked hinges and other 
defects  because  of  heat  expansion.  If  you  wrap 
instruments,  make  sure  your  towels  do  not  contain 
detergent  residue,  which  can  stain  your  instruments. 
Make  sure  that  the  towels  used  in  sterilization  of 
instruments have no detergent residue and are neutral 
pH (7) if immersed in water. This can be a problem, as 
laundries will frequently use inexpensive but high pH (9‐
13)  detergents  and  do  not  properly  rinse  out  or
neutralize  those  detergents  in  the  final  wash  or  rinse 
cycle. Also, sometimes bleaches are added and are not
neutralized.  Hospitals  use  a  "sour"  rinse  cycle  to 
neutralize all detergent residue. 
For Instrument Sets  
Unlock  all  instruments  and  sterilize  them  in  an  open 
position. Place the heavier instruments on the bottom of 
the  set,  when  two  layers  are  required.  Never  lock  an 
instrument during autoclaving. Steam cannot penetrate 
contacting surfaces and this will compromise the sterility 
of the instruments. The instrument will develop cracks in 
hinges  (lock  box)  areas  because  of  the  heat  expansion 
during sterilization. Do not overload autoclave chambers 
as  pockets  may  form  and  will  not  permit  steam  to 
penetrate.  Place  a  towel  on  the  bottom  of  the  pan  to 
absorb the excess moisture during autoclaving. 
CAUTION:  With  most  portable,  tabletop  autoclaves,  at 
the  end  of  the  autoclave  cycle  –  before  drying  cycle  – 
unlock  the  door  and  open  it  no  more  than  a  crack 
(approximately 1/4"). Then run dry cycle for the period 
recommended  by  the  autoclave  manufacturer.  If  the 
autoclave  door  is  opened  fully  before  the  drying  cycle, 
cold  room  air  will  rush  into  the  chamber,  causing 
Canadian Hospital Specialties | 2810 Coventry Road, Oakville, Ontario L6H 6R1  
1‐800‐461‐1423 | [email protected] | www.chsltd.com
 Canadian Hospital Specialties Ltd.  
condensation on the instruments. This will result in water 
stains  on  the  instruments  and  also  cause  wet  packs. 
Make  sure  the  autoclave  filters  and  chambers  are 
cleaned periodically. A stain remover will remove stains 
and clean the autoclave chamber. 
F) Cold Sterilization 
Most  cold  sterilization  solutions  render  instruments 
sterile  only  after  10‐hour  immersions.  This  prolonged 
chemical  action  can  be  more  detrimental  to  surgical 
instruments than the usual 20‐minute autoclave cycle. If 
the  instrument  needs  to  be,  "disinfected"  only,  a  cold 
sterilization soak is okay, as disinfecting will take place in 
10‐minutes  or  more.  Check  the  manufacturer's 
specifications. Also, see the warning sign on using bleach 
(section B). 
Keep in mind the difference between:  
STERILE – No living microorganisms   
DISINFECTED – Kills most but not all microorganisms 
CAUTION: For instruments with Tungsten Carbide inserts 
(hemostats, scissors, needle holders – also identified by 
a gold handle), we do not recommend cold sterilization 
or  solutions  containing  Benzyl  Ammonium  Chloride, 
which will deteriorate the Tungsten Carbide, inserts. 
Rev.IN-INSTRUM.0