Button Functions
ALARM ON/OFF BUTTON
The Biomet
SpinalPak
®
charged battery pack is inserted. The button located below the LCD display enables or disables the
audible alarm. During an alarm condition, depressing the button quickly (0.5 seconds) will temporarily
disable the audible alarm. Depressing the button for a longer period of time (3 seconds) will toggle
the audible alarm between enabled and disabled. Patients should be advised to leave the audible alarm
enabled as frequently as possible in order to assure the fully prescribed treatment. A speaker symbol
will be indicated on the LCD display when the alarm is enabled.
LCD Symbol Descriptions and Instructions
The alarm defaults to audible alarm. Press the button below the display on the front of the stimulator
to silence the alarm. After silenced, the light will continue to flash and the display will indicate the
alarm condition.
Symbol
Non-invasive Spine Fusion Stimulator System is activated as soon as a
®
Condition
Treating
Audible alarm
Low battery charge
Disconnection
System error
Stimulator is
connected to
a PC
End of operation/
Treatment Completion
Instructions
Continue use.
If beeping, depress the button
briefly to silence the alarm. Depress the
button approximately 3 seconds to
engage or disengage the audible alarm.
Insert a charged battery pack.
Confirm that each electrode is properly
applied on the skin. See the electrode
pouch for instructions. Confirm that the
lead wire is attached properly. Replace
the lead wire if necessary.
Error in the stimulator –
Contact Biomet for assistance.
Stimulator will not treat until USB
cable is disconnected
Contact Biomet
9