Recommended separation distances between portable and
mobile RF communications equipment and the Biomet
SpinalPak
The Biomet
SpinalPak
®
electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user
of the Biomet
SpinalPak
®
interference by maintaining a minimum distance between portable and mobile communications equipment
(transmitters) and the Biomet
below, according to the maximum power output of the communications equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1. At 80 MHz and 800 MHz, the separation distance for the higher frequency applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Non-invasive Spine Fusion Stimulator System
®
Non-invasive Spine Fusion Stimulator System is intended for use in an
®
Non-invasive Spine Fusion Stimulator System can help prevent electromagnetic
®
SpinalPak
Non-invasive Spine Fusion Stimulator System as recommended
®
®
according to frequency of transmitter
150 kHz to 80
MHz
d = 3.5 √ P
.35
1.1
3.5
11.06
35
Table 4
Separation distance (meters)
80 MHz to 800
MHz
d = 3.5 √ P
.35
1.1
3.5
11.06
35
15
®
800 MHz to 2.5
GHz
d = 7 √ P
.7
2.21
7
22.13
70