Dickey-John GAC 2500 UGMA 매뉴얼 - 페이지 2
{카테고리_이름} Dickey-John GAC 2500 UGMA에 대한 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Dickey-John GAC 2500 UGMA 12 페이지. Grain analysis computer
ADDENDUM (GAC® 2500 UGMA/INTL)
/
2
MAINTENANCE
INTERNAL CLEANING
Performing continuous tests can result in material accumulation around the
critical internal components and adversely affect the measurement.
Two types of cleaning are recommended on an as needed basis:
•
Daily clean
•
Extensive clean
DAILY CLEAN METHOD
A daily clean method allows cleaning the cell and door using an automated
process. During the cleaning sequence, the hopper door automatically
opens. Instruments with version 2.4 grain handler firmware, the trap door
opens automatically also.
To start the cleaning process:
1. At the Main Menu screen, press the Instrument Information button.
2. At the Instrument Information screen, press the Clean button.
– Pressing the Clean button automatically begins the cleaning
sequence and opens the hopper door and the trap door (version 2.4
grain handler firmware).
Figure 2
Daily Clean Method
DICKEY-john Corporation
-
Analyze
Cleaning Cell
Setting Instrument to Cleaning Mode
GAC2500-UGMA
Current User: Guest
Results
Setup
Instrument Information
Model: GAC2500
S/N: 1807-00005
Software Versions:
4.2.0.0, 2.4, 2.2
IP Address: 169.254.0.37
Last Service: 1/1/2014
10-038
NTEP COC:
Directive No: 9180.61 05-01-13
GAC2500-UGMA
Please wait...
GAC® 2500 Models (UGMA & INTL)
GAC2500-UGMA
11001-1646A-201409