Dickey-John GAC 2500 UGMA Podręcznik - Strona 2

Przeglądaj online lub pobierz pdf Podręcznik dla Sprzęt rolniczy Dickey-John GAC 2500 UGMA. Dickey-John GAC 2500 UGMA 12 stron. Grain analysis computer

ADDENDUM (GAC® 2500 UGMA/INTL)
/
2
MAINTENANCE

INTERNAL CLEANING

Performing continuous tests can result in material accumulation around the
critical internal components and adversely affect the measurement.
Two types of cleaning are recommended on an as needed basis:
Daily clean
Extensive clean

DAILY CLEAN METHOD

A daily clean method allows cleaning the cell and door using an automated
process. During the cleaning sequence, the hopper door automatically
opens. Instruments with version 2.4 grain handler firmware, the trap door
opens automatically also.
To start the cleaning process:
1. At the Main Menu screen, press the Instrument Information button.
2. At the Instrument Information screen, press the Clean button.
– Pressing the Clean button automatically begins the cleaning
sequence and opens the hopper door and the trap door (version 2.4
grain handler firmware).
Figure 2
Daily Clean Method
DICKEY-john Corporation
-
Analyze
Cleaning Cell
Setting Instrument to Cleaning Mode
GAC2500-UGMA
Current User: Guest
Results
Setup
Instrument Information
Model: GAC2500
S/N: 1807-00005
Software Versions:
4.2.0.0, 2.4, 2.2
IP Address: 169.254.0.37
Last Service: 1/1/2014
10-038
NTEP COC:
Directive No: 9180.61 05-01-13
GAC2500-UGMA
Please wait...
GAC® 2500 Models (UGMA & INTL)
GAC2500-UGMA
11001-1646A-201409