When using IX83, the auto compensation of the stage can be performed
from the touch panel controller.
Before starting the auto compensation, install the stage on the
IX83 and make sure nothing, including a specimen, is placed
on the stage.
1. Open the [Menu] display by controlling the touch panel.
2. Tap the System Setting button @ to open the [System Setting] display.
3. Tap the [System] tab 2 and then tap the Auto Compensation
button 3.
4. When the auto compensation confirmation display appears, tap [OK] to
start auto compensation.
The stage moves automatically during the auto compensation
(for a few minutes). Do not touch the stage during the movement.
5. When auto compensation completes, the touch panel screen shows the
completion message.
If the foreign substance is entered in the stage driving part,
the status may not be recovered by performing the Auto
Compensation. In this case, turn the power of BX3-SSU Control
Box to "
" (OFF), and perform the Auto Compensation again
after moving the stage from front to back and from side to side
several times with a hand.
1. Set the power of the control box for the scanning stage with ultrasonic
to "
" (OFF).
2. Move the stage to the origin position with hands.
Horizontal direction: Align the edge on the upper level to the middle level
of the stage.
Vertcal direction: Align the guide part 1 placed at the bottom of the middle
level of the stage to the edge of the front side of the lower level.
3. Set the power of the control box for the scanning stage with ultrasonic
to " I " (ON).
If the power of the control box for the scanning stage with
ultrasonic is set to " I " (ON) while the stage is not placed at the
origin position, the origin position is misrecognized to limit the
stage movement range.
BX3-SSU
8