Olympus BXC-CBB 사용 설명서 - 페이지 14

{카테고리_이름} Olympus BXC-CBB에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus BXC-CBB 41 페이지. Modular microscope assemblies hardware

3.5 Setting up BXC-CBB

TIP:
Before setting the DIP switch, set the main switch to ○(OFF).
The switch settings are read and defined after Power-ON.
1.
Confirm that the DIP switch on BXC-CBB is set properly.
The allocated functions of the DIP switches are shown in the table below.
: Factory default setting
SW-No.
1
2
3
4
5
Off
On
Off
On
Off
On
Off
On
Off
On
3. Setup procedures
Function
6
7
8
Buzzer sound
Nosepiece (NP)
1
AS control*
NP Control*
Reserved by maker Always OFF
Off
Reserved by maker Always OFF
On
Off
Reserved by maker Always OFF
On
Off Reserved by maker Always OFF
On
The buzzer sound is heard
The buzzer sound is not heard
5Position-NP
6Position-NP
Normal
High-precision control (lost motion reduction)
2
Rotate through the highest numbered hole
Does not rotate through the highest numbered hole
Note
3.5 Setting up BXC-CBB
10