Olympus CK40M FRAME 수리 매뉴얼 - 페이지 36

{카테고리_이름} Olympus CK40M FRAME에 대한 수리 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus CK40M FRAME 38 페이지. Compact inverted metallurgical microscope

Olympus CK40M FRAME 수리 매뉴얼
CK40M
9.

Fine focus sensitivity check

9-1
Set one of following products as illustrated below to check the fine focus sensitivity.
• Eyepiece--------------- WHK10X, WHK15X, NCWHK10X, GSWHK10X
• Objective -------------- MDPlan5X, MDPlan10X, MDPlan20X, MDPlan40X,
• Observation tube ----- CH3-BI45, CH3-TR45, CK40-TBI
• Specimen-------------- Metallic specimen
9-2
Checking fine focus sensitivity
1)
Focus on the metallic specimen.
2)
Remember the position of the scale on the fine adjustment knob at the focused
position.
3)
Turning the fine adjustment knob, set the specimen out of focus (in the both
directions), and then focus on the specimen again.
4)
If the position of the scale on the fine adjustment knob is within 2 graduations from
the position in 2), the result is acceptable.
5)
If the displacement exceeds 2 graduations, the focusing unit must be assembled
again.
Standard:
C. REPAIR PROCEDURE
MDPlan50X, MDPlan80X
MDAch10X, MDAch20X, MDAch50X
(CH3-BI45 is mounted in the following illustration.)
Metallic specimen
Within 2 graduations from the previous focus position
Observation tube
C-15
Eyepiece
Objective