Olympus MX61L 조립 설명서 - 페이지 35

{카테고리_이름} Olympus MX61L에 대한 조립 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus MX61L 49 페이지. 200mm/300mm compatible semiconductor/fpd inspection microscopes

Olympus MX61L 조립 설명서

Objective Switching

Functions
1. Rotates the objective in the specified revolving nosepiece position (objective) in the optical axis.
2. Checks the current revolving nosepiece position.
Special notes
1. If request command 1OB
2. If the revolving nosepiece position is not in the optical axis or the sensor is abnormal, "X" is returned as the response in
place of a number between "1" and "6".

Command descriptions

Command
Type
1OB
n
Request
(n: 1 to 5 when a
5-position nose-
piece is used, or
1 to 6 when a 6-
position nose-
piece is used)
1OB?
Query
*
)
"
" represents a space.
Line feed command <CR><LF> is required after commands and line feed command <CR><LF> are added after any
response.
6 is sent while a 5-position revolving nosepiece is used, an illegal parameter error will occur.
Description
Response
Rotation of the re-
1OB
volving
nose-
piece position
(objective) with
1OB
the
number
specified with "n"
1OB
in the optical axis.
1OB
Check of the
1OB
current revolving
(n: 1 to 5 or 6.
nosepiece posi-
X = Undefined.)
tion (objective)
1OB
number.
Result
+
Successful
!,E11214
Error
!,E01130
Error
!,E01120
Error
n
Successful
!,E01150
Error
MX61/MX61L
1OB 1OB?
Description
The specified revolving nose-
piece position is engaged in the
optical axis.
It cannot be rotated in the opti-
cal axis due to a time-out error.
The nosepiece is not connected
or the connected nosepiece is
not the designated model.
A string other than "1" to "6" is
sent as "n".
The nosepiece position that is
currently engaged in the optical
axis is returned.
A string is sent after "1OB?
".
32