Olympus CKX41 사용 설명서 - 페이지 25
{카테고리_이름} Olympus CKX41에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus CKX41 36 페이지. Reflected fluorescence system
Olympus CKX41에 대해서도 마찬가지입니다: 개요 (7 페이지), 브로셔 및 사양 (4 페이지), 수리 매뉴얼 (43 페이지)
- 1. Table of Contents
- 1. General Precautions for Observation
- 1. Using the Controls
- 2. Important
- 2. Applicable Fluorescence Objectives
- 3. Nomenclature
- 3. Centering the Mercury Burner
- 4. Troubleshooting Guide
- 4. Specifications
- 4. Assembly
- 4. Assembly
- 4. U-Excitation Filter Set Ckx-Nu (Optional)
- 4. Controls of each Module
- 4. Switching the Filter Slider
- 5. Proper Selection of the Power Supply Cord
- 5. Summary of Reflected Fluorescence Observation Procedure
7-2 Detailed Assembly Procedure
2
3
Fig. 10
1
2
Fig. 11
3
4
Fig. 12
1
Removing the Dummy Cover
1. Rotate the coarse adjustment knob 1 in the direction of the
arrow to raise the revolving nosepiece to the upper limit.
2. Hook you finger into the pulling hole 3 on the cover 2 and
remove the cover.
1
Removing the Dust Cover and
2
Fluorescent Mirror Slider
1. Using the Allen wrench provided with the illuminator, loosen
2
the clamping knob 2 of the dust cover 1 and remove it.
2. Using the Allen wrench provided with the illuminator, loosen
the stopper screw 4 of the fluorescent mirror slider 3 at the
extremity of the CKX-RFA illuminator and remove the screw
and washer.
3. Pull the fluorescent mirror slider 3 in the direction of the
arrow to remove it.
REFLECTED FLUORESCENCE SYSTEM
FOR CKX41
(Fig. 10)
(Figs. 11 & 12)
22