Olympus SZ2-ILLC 사용 설명서 - 페이지 15
{카테고리_이름} Olympus SZ2-ILLC에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus SZ2-ILLC 16 페이지. Coaxial illuminator
Olympus SZ2-ILLC에 대해서도 마찬가지입니다: 사용 설명서 (16 페이지)
5 5
Troubleshooting Guide
Under certain conditions, performance of the unit may be adversely affected by factors other than defects. If problems
occur, please review the following list and take remedial action as needed. If you cannot solve the problem after
checking the entire list, please contact Olympus.
Problem
Field of view is not evenly
The light guide is attached improperly.
illuminated.
The lamp bulb is attached improperly
Color reproduction in the
The light intensity control knob of the
photomicrographs is poor.
light source is not set to the maximum
(LG-PS2 only)
position.
Filters (30.5S-LB80 and 30.5S-LBD) are
not mounted (when a daylight type film
is used).
Cause
Attach it properly.
Mount it properly.
Set the light intensity control knob to the
maximum position.
Mount the filters (30.5S-LB80 and
30.5S-LBD) on the LG-PS2 light source.
SZ2-ILLC
Remedy
Page
10
–
4
–
12