4.2. Prepare the Drill
Nonsterile Team Member
1. Install a charged battery and confirm that
the drill is functioning by pressing the
trigger and observing if it turns ON. If the
drill is not functioning, do not proceed to
the next step. Try re-charging the battery
again or use a different charged battery.
If the drill still does not operate, contact
Arbutus Medical.
2. You may terminate operation of the drill by
releasing the trigger.
3. Ensure toggle switch is set to forward, not
reverse.
4. Set the torque to "MAX"
A torque setting less than 'MAX' may be
insufficient for drilling.
4.3. Prepare Sterile Cover
Sterile Team Member
1. Hold Cover and keep it suspended in air.
2. Roll the Cover edge back. Only expose the
inside lining of the Cover to the non-sterile
team member.
3. Hold the Cover so that your sterile gloves
are tucked under the folded fabric cuff and
protected by the fold.
Arbutus Medical