Siemens sitrans 사용 설명서

Installation

Ideal Locations

areas where the acoustic emission levels are highest and most consistent
where the material impact is greatest
areas closest to the point of leakage (e.g. valve body)
closest point to source of cavitation (e.g. pump body)

Conditions to Avoid

non-metallic surfaces, as these tend to attenuate acoustic emission levels
pipes with non-metallic liners as these tend to attenuate acoustic emission levels

Conditions to Consider

joints and interfaces attenuate the acoustic emission levels.
minimal temperature variation where acoustic emission levels are weak.
location should provide sufficient response time in alarm or control circuit. e.g. loss
of bearing lubrication, plug chute detection.
Solids Flow Sensing
Recommended
Good friction between product and
piping. Mount Sensor on the
upstream side of gasket for highest
level of acoustic emission activity.
Page 6
SITRANS AS 100 – INSTRUCTION MANUAL
Not Recommended
Minimal amount of acoustic emission
activity due to limited friction
between product and piping.
7ML19985DM02