Siemens ULTRAMAT 6F 사용 설명서 - 페이지 44

{카테고리_이름} Siemens ULTRAMAT 6F에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Siemens ULTRAMAT 6F 49 페이지. Gas analyzer for the determination of ir-absorbing gases
Siemens ULTRAMAT 6F에 대해서도 마찬가지입니다: 추가 장비 매뉴얼 (20 페이지), 안전 매뉴얼 (24 페이지)

Siemens ULTRAMAT 6F 사용 설명서
Sample gas inlet conditions
Sample gas inlet conditions of analyzer versions
Type of gas
Max. pressure p
max
Min. pressure p
min
Temperature
Sample gas flow rate q
v
44
Non-combustible gases or gas mixtures that
always remain below the LEL in accordance
with Category 4 from EN ISO 13849-1: 2016
1100 hPa for CALOMAT 6F and
CALOMAT 62F
3000 hPa for OXYMAT 6F
1500 hPa for ULTRAMAT 6F
500 hPa for OXYMAT 6F
600 hPa for ULTRAMAT 6F
800 hPa for CALOMAT 6F and
CALOMAT 62F
CALOMAT 6F: 0 to 50 °C
CALOMAT 62F: 0 to 50 °C
OXYMAT 6F: 0 to 50 °C
OXYMAT 6F: In heated version: 0 to 130 °C, but only 15 °C above temperature analyzer unit
ULTRAMAT 6F: 0 to 50 °C
ULTRAMAT 6F: In heated version: 0 to 80 °C
CALOMAT 6F / CALOMAT 62F: 0.5 ... 1.5 l/min
OXYMAT 6F: 0.3 ... 1.0 l/min
ULTRAMAT 6F: 0.3 ... 1.5 l/min
Combustible gases up to gas mixtures which
are occasionally explosive
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
For combustible gases and analyzer ver-
sions Ex px with "leakage compensation"
100 hPa (rel.) for ULTRAMAT 6F and
OXYMAT 6F
Analyzer versions Ex px with "continuous
purging" as well as
1100 hPa for CALOMAT 6F and
CALOMAT 62F
3000 hPa for OXYMAT 6F
1500 hPa for ULTRAMAT 6F
For combustible gases and analyzer ver-
sions Ex px with "leakage compensation"
-200 hPa (rel.) for ULTRAMAT 6F and
OXYMAT 6F
A5E45779389002-01, 09/2018