Siemens UROSKOP D 추가 설치, 작동 및 유지보수 지침 - 페이지 10

{카테고리_이름} Siemens UROSKOP D에 대한 추가 설치, 작동 및 유지보수 지침을 온라인으로 검색하거나 PDF를 다운로드하세요. Siemens UROSKOP D 26 페이지.

Siemens UROSKOP D 추가 설치, 작동 및 유지보수 지침
2 - 2
Fig. 1
2.2.3

Lifting drive

SIM
Actuator cams
SIE
Safety limit switch
Slide the actuator cams (1/Fig. 1) of safety switches s21 and s22 by hand (when
activating the switch, the K1 contactor in M20 drops) At the same time, check that the
actuator cam moves easily.
Raise the system to +88°.
SIE
Travel range
Move the system vertically over the complete range of travel, checking for noise and smooth travel.
Measure the maximum vertical position 1080 mm and the minimum vertical position 620
mm with a tape measure between the lower edge of the lifting base and the rotating axis.
SIM
Chain coupler
Check the chain coupler position on the outside of the lifting base (2/Fig. 1/2); the bolt for
the chain coupler (K1) must firmly connect with the outer surface of the lifting base (F)
(Fig. 2).
Check that the safety screw (S) is secured tightly.
UROSKOP D
Inspection and maintenance
Fig. 2
RLL5-310.101.02
Page 2 of 16
Rev. 02
01.97
TDSP 1
Siemens AG
Medical Engineering