Siemens UROSKOP D Ek Kurulum, Çalıştırma ve Bakım Talimatları - Sayfa 10
Endüstriyel Ekipmanlar Siemens UROSKOP D için çevrimiçi göz atın veya pdf Ek Kurulum, Çalıştırma ve Bakım Talimatları indirin. Siemens UROSKOP D 26 sayfaları.
2 - 2
Fig. 1
2.2.3
Lifting drive
SIM
Actuator cams
SIE
Safety limit switch
•
Slide the actuator cams (1/Fig. 1) of safety switches s21 and s22 by hand (when
activating the switch, the K1 contactor in M20 drops) At the same time, check that the
actuator cam moves easily.
•
Raise the system to +88°.
SIE
Travel range
•
Move the system vertically over the complete range of travel, checking for noise and smooth travel.
•
Measure the maximum vertical position 1080 mm and the minimum vertical position 620
mm with a tape measure between the lower edge of the lifting base and the rotating axis.
SIM
Chain coupler
•
Check the chain coupler position on the outside of the lifting base (2/Fig. 1/2); the bolt for
the chain coupler (K1) must firmly connect with the outer surface of the lifting base (F)
(Fig. 2).
•
Check that the safety screw (S) is secured tightly.
UROSKOP D
Inspection and maintenance
Fig. 2
RLL5-310.101.02
Page 2 of 16
Rev. 02
01.97
TDSP 1
Siemens AG
Medical Engineering