Adjustments
4.2
Calibrating the Diameter Recognition
Switch on the system.
Start menu.
Setup >
Select
Insert the calibration rod 10 mm (Part No. 5988543).
Select Continue.
The wrapping unit will be turned ten times with closing and opening the jaws and measuring the diameter of the
test pipe.
The display shows the success of the calibration, the determined offset and the tolerances of the offset value.
Select Continue.
The determined offset will automatically be stored and considered for future measurements.
Labelling >
Calibrate device.