Adjustments
4.2

Calibrating the Diameter Recognition

 Switch on the system.
 Start menu.
Setup >
 Select
 Insert the calibration rod 10 mm (Part No. 5988543).
 Select Continue.
The wrapping unit will be turned ten times with closing and opening the jaws and measuring the diameter of the
test pipe.
 The display shows the success of the calibration, the determined offset and the tolerances of the offset value.
 Select Continue.
The determined offset will automatically be stored and considered for future measurements.
Labelling >
Calibrate device.
15