{카테고리_이름} Dinel RFLS-53N에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Dinel RFLS-53N 16 페이지. High frequency level switch
Setting modes
The sensor can be set to normally closed "O-mode" or to normally open "C-mode" switch types.
level state
level state
For safety reasons, for scanning min. level, we recommend you to use the "O" mode setting (sensor closes when
submerged). A faulty sensor or wiring will take effect here in the same way as level emergency conditions by opening
the sensor. Analogously for the max. level, we recommend you to set the "C" mode (sensor opens when submerged).
© Dinel, s.r.o.
RFLS–53
mode
output state
O
CLOSED
O
OPEN
mode
output state
C
OPEN
C
CLOSED
Orange LED
(shining)
(dark)
Orange LED
(dark)
(shining)
11