configure or operate the CoriolisMaster meters, refer to the CoriolisMaster FCB400, FCH400 Operation
Instructions (See
Additional information
Pre-requisites:
The CoriolisMaster is powered.
The CoriolisMaster configuration is complete (Digital Output 41/42 is configured as required).
Use the operator buttons to navigate on the CoriolisMaster local display:
1.
Press the left button to navigate to the Operator Menu.
2.
Navigate and select Diagnostics.
Figure 7-1: CoriolisMaster Diagnostics on the LCD
3.
Scroll down, locate, and select Output Readings.
4.
Scroll down and locate the DO 41/42 Freq (Dig.Out 41/42 Freq.) parameter.
5.
Record the displayed value.
1000 Hz.
Figure 7-2: ABB CoriolisMaster DO 41/42 frequency output
8

Verify input detection at the HSPC module

This procedure verifies that the HSPC module receives pulse inputs from the CoriolisMaster.
Prerequisites:
HSPC module and CoriolisMaster are connected and powered.
To verify pulse detection by the module:
On the PCCU navigation tree, expand ONLINE, then select TFIO 8/HSPC I:00.
1.
Select the Pulse Counts tab.
2.
3.
Verify that the PI 1 Pulse Count matches the frequency output displayed by the CoriolisMaster
LCD.
for a link).
Figure 7-2
shows an example of a DO 41/42 frequency output of
21 08 74 4MN A A | HS P C C ON FIG UR AT IO N G UI DE | 9