Alcohol Countermeasure Systems ALCOSCAN 캘리브레이션 매뉴얼 - 페이지 6

{카테고리_이름} Alcohol Countermeasure Systems ALCOSCAN에 대한 캘리브레이션 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Alcohol Countermeasure Systems ALCOSCAN 16 페이지. Passive breath alcohol tester

SAFETY AND PRECAUTIONS

WARNING! READ CAREFULLY before beginning the calibration process.

GENERAL

This guide is intended for authorized service technicians only
Do not use the ALCOSIM simulator with toxic liquids, flammable liquids or in
explosive atmospheres
Use the product for its intended purpose only; failure to do so will void the warranty,
may cause injury and damage the components
CAUTION – Hot Surface! Avoid contact with the ALCOSIM simulator heating
element
Only use parts supplied by Alcohol Countermeasure Systems Corp
Ensure that the proper amount of solution is added; do not under-fill or over-fill the
ALCOSIM simulator
If the ALCOSIM simulator overheats considerably beyond 34 °C, turn it off and
contact an authorized service provider
Before use, inspect all components for visible cracks or damage; failure to do so
may cause injury or additional damage
Never run the ALCOSIM simulator without adding alcohol reference solution
Perform calibration within a room temperature environment (23 °C ± 2 °C and
approx. 50% RH)
DO NOT overtighten the solution container

MOISTURE, CONTAMINATION AND CLEANING

Always disconnect the ALCOSIM simulator power before cleaning
If the simulator was on: before removing the top housing, ensure that the simulator
is switched off, unplugged AND allow at least 15 minutes for the heater to cool
down
Remove the ALCOSIM simulator top housing before cleaning the solution container
Wash the solution container with plain water and dry with a paper towel
Clean the top housing by wiping with a damp cloth; DO NOT submerge the top
housing in water
Before setup, ensure that all moisture has been completely removed from the
ALCOSIM simulator
Before setup, all tubing and mouthpieces must be completely dry. Even slight
condensation may disrupt calibration
Completely empty the ALCOSIM solution container at the end of the work day
1
ALCOSCAN