Alcohol Countermeasure Systems ALCOSCAN Instrukcja kalibracji - Strona 6
Przeglądaj online lub pobierz pdf Instrukcja kalibracji dla Sprzęt testowy Alcohol Countermeasure Systems ALCOSCAN. Alcohol Countermeasure Systems ALCOSCAN 16 stron. Passive breath alcohol tester
SAFETY AND PRECAUTIONS
WARNING! READ CAREFULLY before beginning the calibration process.
GENERAL
•
This guide is intended for authorized service technicians only
•
Do not use the ALCOSIM simulator with toxic liquids, flammable liquids or in
explosive atmospheres
•
Use the product for its intended purpose only; failure to do so will void the warranty,
may cause injury and damage the components
•
CAUTION – Hot Surface! Avoid contact with the ALCOSIM simulator heating
element
•
Only use parts supplied by Alcohol Countermeasure Systems Corp
•
Ensure that the proper amount of solution is added; do not under-fill or over-fill the
ALCOSIM simulator
•
If the ALCOSIM simulator overheats considerably beyond 34 °C, turn it off and
contact an authorized service provider
•
Before use, inspect all components for visible cracks or damage; failure to do so
may cause injury or additional damage
•
Never run the ALCOSIM simulator without adding alcohol reference solution
•
Perform calibration within a room temperature environment (23 °C ± 2 °C and
approx. 50% RH)
•
DO NOT overtighten the solution container
MOISTURE, CONTAMINATION AND CLEANING
•
Always disconnect the ALCOSIM simulator power before cleaning
•
If the simulator was on: before removing the top housing, ensure that the simulator
is switched off, unplugged AND allow at least 15 minutes for the heater to cool
down
•
Remove the ALCOSIM simulator top housing before cleaning the solution container
•
Wash the solution container with plain water and dry with a paper towel
•
Clean the top housing by wiping with a damp cloth; DO NOT submerge the top
housing in water
•
Before setup, ensure that all moisture has been completely removed from the
ALCOSIM simulator
•
Before setup, all tubing and mouthpieces must be completely dry. Even slight
condensation may disrupt calibration
•
Completely empty the ALCOSIM solution container at the end of the work day
1
ALCOSCAN