4. Plug the power cord into an electrical outlet with the appropriate voltage.
NOTE: The pump unit must be suitable for the local power voltage.
5. Turn the main power switch on pump to the ON position. After you connect power to the
pump, you will have to wait at least 30 minutes for the mattress to become fully inflated. The
low-pressure indicator (yellow or red LED) will illuminate while the mattress is pressurizing. The
audible alarm won't sound within 40 minutes; after that, the audible alarm will be activated if the
pressure is still low inside the mattress.
When the appropriate pressure is reached, the low-pressure indicator (yellow or red LED) will go off
and the normal-pressure indicator (green LED) will illuminate. The bed is now ready for use.
NOTE: You can unplug the unit to power off the device.
•
CAUTION: The pump supplied with the mattress
can only be used for mattresses recommended by the
manufacturer. Do not use it for any other purpose.
NOTE: (For models WITHOUT the low air loss function)
During power outages, you can cover the Quick
Connector with the transport cap to maintain air pressure
inside the air cells.
NOTE: After installation, make sure any excess cord
C
or tubing is removed from possible foot traffic to avoid
P
accidental tripping. All equipment should be positioned to
R
always allow unhindered resident access by physicians
and caregivers.
1-800-634-7328
DirectSupply.com
7