TEST PROGRAM
©TEST-OD : Alignment offset adjustment
Press this
key to quit
TEST-OD.
• Adjusting the alignment signal offsets (with no room lamp)
"T": Offset of the TOT signal caused by the amplifier system
Adjust VR6 TOT to set " R ± 10".
"X": Offset of the XAL signal caused by the amplifier system
Adjust VR5 XAL to set " R ±10 ".
" Y": Offset of the YAL signal caused by the amplifier system
Adjust VR4 YAL to set " R ±10 ".
"R": Reference voltage level for the alignment signal (2.5V)
Make sure that the reference voltage is
between 120 and 135 (not adjustable).
• Adjusting the peak value of CHA - FIL and CHB - FIL.
"A": Peak value of the CHA - FIL signal when the C LED is on.
Make sure that the peak value is between 150 and 200.
"B": Peak value of the CHB - FIL signal when the C LED is on.
Make sure that the peak value is between 150 and 200.
* To adjust "A" and "B" use VR7 CAL to measure the peak values for both
channels so that they fall within the permitted value range (same as in
TEST-06).
• 0V display
"0": Make sure that "000" is displayed.
[6-10]
Repair Manual