V-5 USER MANUAL
7.3.2
Device Error Menu Map
Errors
Over heat shutdown
Heat up failed
Pump protection
Thermocouple failed
Thermocouple
tripped off
No remaining run-
time
Gas tank
pressurization failed
Tank pressure high
Display
Possible Cause
Device Error
When the temperature is higher than the limit temperature, the over
temperature protection function is activated.
E0
Device Error
Heating system cannot reach pre-set target temperature after 10-min
heat-up
E1
Device Error
Working current of the air pump exceeds the preset value, the error
message will be activated.
E2
Device Error
Short circuit or wrong connection of thermocouple
E3
Device Error
Open circuit or no connection of thermocouple
E4
Device Error
No remaining run-time
E5
When the air pump pressurization function is turned on, if it cannot be
Device Error
pressurized normally within 2 minutes, the error message will be
E6
activated.
Device Error
When the pressure of the gas tank is greater than 60, the error
message will be activated.
E7
14