Ashcroft GC52 제품 정보 페이지 - 페이지 8

{카테고리_이름} Ashcroft GC52에 대한 제품 정보 페이지을 온라인으로 검색하거나 PDF를 다운로드하세요. Ashcroft GC52 8 페이지. Differencial pressure transmitter

Product Information Page
Analog output check:
Similarly, analog output can be verified by placing an amp-meter in series with the transmitter output
and power supply (see manual wiring instructions for reference). As well, during measurement mode or
loop check (see results below for reference).
Connect amp meter per manual
instructions or figure above.
Press the Down arrow until 0.0 GPM is
displayed or apply 0 IWC.
Verify amp meter reading (4.00 mA).
0.0 GPM of water corresponds to 0% FS (4
mA at 0.0 GPM of water).
Press the Up or Down arrow until 44.0 GPM
is displayed or apply 10 IWC.
Verify amp meter reading (12.00 mA).
44.0 GPM of water corresponds to 50% FS
(12 mA at 44.0 GPM of water).
Press the Up arrow until 88.0 GPM is
displayed or apply 20.0 IWC.
Verify amp meter reading (20.0 mA).
88.0 GPM of water corresponds to 100% FS
(20 mA at 88.0 GPM of water).
All specifications are subject to change without notice.
All sales subject to standard terms and conditions.
© 2017 Ashcroft Inc.
ashcroft.com
info@ ashcroft.com
1.800.328.8258
Rev. A 02/17, Page 8 of 8