AMC AMC-1225 Instructions Manual - Page 15
Browse online or download pdf Instructions Manual for Security Sensors AMC AMC-1225. AMC AMC-1225 17 pages. Sensor modules
AMC-1225 SENSOR MODULES
NOTE:
If the sensor has been replaced, adjust the SIGNAL first, then proceed with LOW alarm
and HIGH alarm adjustments. Signal calibration is performed using a digital multimeter
set to measure DC voltages to two (2) decimal places (e.g. 0.00 Volts).
The SIGNAL adjust is used to set the sensitivity of the sensor. This is done by exposing the
sensor to the LOW alarm concentration of gas. The voltage associated with signal can be
measured between the negative (-) power line on the output terminal block and the test point
Figure 5-1Figure 5-1
next to the signal trimmer, see
. The voltage measured at the signal test
point should be approximately 1.20 Volts.
When calibration is completed, remove the calibration chamber, replace the sensor element
while observing the correct polarity, and then reinstall the cover plate with the four screws.
CAUTION:
Only qualified personnel should perform the actual calibration. Users are advised to
consult The Armstrong Monitoring Corporation as to the recommended calibration gas
concentration for the application, and any other questions.
14