2. Front panel description

Alarm outputs indicators:
»
4
Measurement chain diagnostics for IEPE sensor:
»
Res
LEDs:
open - open circuit or sensor failure
short - short circuit or sensor failure
Indicator of the selected estimate:
»
Green LEDs:
RMS – RMS value of vibration signal
PEAK – maximum value of vibration signal (0-Peak)
Indicator of the selected measured value:
»
Green LEDs:
acc – acceleration
vel – velocity
Red LED:
ALARM – the alarm threshold exceeded, alarm output
ON
Yellow LED:
WARNING – the warning threshold exceeded, warning
ON
Proper work indicator:
»
Green LED:
OK - diode pulsing with frequency of 1 Hz indicates the
correct operation of the device
rapid pulsing of OK diode means entering the device
setup menu
Keyboard:
»
UP – up
DN - down
SET - set
Measured value:
»
double-digit LED display
AV MONITOR 1002D