amc VIBRO AVM 1002D Manual do utilizador - Página 4
Procurar online ou descarregar pdf Manual do utilizador para Instrumentos de medição amc VIBRO AVM 1002D. amc VIBRO AVM 1002D 14 páginas. Vibration monitoring system with relay outputs
2. Front panel description
Alarm outputs indicators:
»
4
Measurement chain diagnostics for IEPE sensor:
»
Res
LEDs:
open - open circuit or sensor failure
short - short circuit or sensor failure
Indicator of the selected estimate:
»
Green LEDs:
RMS – RMS value of vibration signal
PEAK – maximum value of vibration signal (0-Peak)
Indicator of the selected measured value:
»
Green LEDs:
acc – acceleration
vel – velocity
Red LED:
ALARM – the alarm threshold exceeded, alarm output
ON
Yellow LED:
WARNING – the warning threshold exceeded, warning
ON
Proper work indicator:
»
Green LED:
OK - diode pulsing with frequency of 1 Hz indicates the
correct operation of the device
rapid pulsing of OK diode means entering the device
setup menu
Keyboard:
»
UP – up
DN - down
SET - set
Measured value:
»
double-digit LED display
AV MONITOR 1002D