8.2. Temperature Calibration
By default ATC mode is enabled.
Temperature calibration is only possible when an ATC probe is fitted. The ATC calibration offset is
limited to ±10 °C.
902857-A - C800 Analyser Operating Manual.docx
Press
again to confirm the slope and the
offset. The analyser will then ask if you would
like to reset the runtime, select NO for this.
The sensor is now calibrated.
To calibrate the temperature, navigate to ATC
in the main menu.
ATC must be enabled.
Place the sensor into a solution of stable
temperature. Adjust the offset, with the
arrows, so that the temperature reads the
correct value and press
The analyser is now calibrated for automatic
temperature readings.
to confirm.