Cole Parmer 1200 Series User Manual - Page 8

Browse online or download pdf User Manual for Measuring Instruments Cole Parmer 1200 Series. Cole Parmer 1200 Series 18 pages.

Cole Parmer 1200 Series User Manual
At F mode, set the factor number and change to C mode (refer to Factor Mode in Basic
Operation section);
If you want to print the test results at C mode, push the MODE button to the position
when the LED light is off, then push the ENT button;
If connected with a PC using UNICO
communication with the PC will be set up (refer to the UNICO
Spectrophotometer Software User's Manual Version SS-1.22 for details).
Sample Compartment: Accept 10 mm test tube or 10 mm square cuvette (with
different cuvette holders). The compartment can accept up to 50 mm pathlength
cuvettes with optional holders.
WAVELENGTH (Wavelength Control) Knob: Select desired wavelength in nanometers
(nm).
Wavelength Readout Window: Display desired wavelength.
Changing Cuvette Holders
There are five different Cuvette Holders (Sample Holders) (refer to Figure-4 from left
to right):
• S-1200-101: Test Tube V-type Cuvette Holder with base for 8~25 mm Diameter
• S-1200-102: Single Square Cuvette Holder with base for 10 mm Cuvettes
• S-1200-103: 4-poaition Cuvette Holder for 10 mm Square Cuvette
• S-1200-104: Longpath 4-position Cuvette Holder for up to 50 mm rectangular
• S-1200-105: Water-jacketed Cuvette Holder with base and panel for
S-1200-101, S-1200-102, and S-1200-105 all consist of two parts: the upper holder and
the identical universal base (S-1200-106: Universal base for V-type holder, Single
Cuvette Holder and water-jacketed Holder)
position Cuvette Holder (Figure-5) with a V-type Tube Holder (Test Tube Holder).
The only tool needed is a standard Philips head screwdriver.
Figure-4
®
Windows-based Application Software, the
Test Tubes
Cuvettes
temperature controlled application
Sample Holders for Model 1200/1201
Follow the steps below to replace a 4-
.
8
®
1200 Series