COM-power corporation T-Series Instruction Manual - Page 8

Browse online or download pdf Instruction Manual for Test Equipment COM-power corporation T-Series. COM-power corporation T-Series 12 pages. Shielded coupling/decoupling networks

5.3
Calibration Accessories
 
 
The following accessories are available for the CDN verification of the parameters above and 
for the level setting calibration described herein. 
 
Shorting adapters (Part ADA‐T4SE / ADA‐T8SE)  
150Ω – 50Ω adapters (Part# ADA‐515)  
50Ω terminator (Part# TEP‐050)  
6.0  Calibrating the Test Levels and Running the Test 
 
6.1   Test Levels 
 
Table 1: Required test levels per IEC‐61000‐4‐6 
V
 
o
Volts 
 
V
 = Open circuit voltage in 150Ω system 
o
Umr= Voltage level measured at the output of the 150Ω to 50Ω adapter (ADA‐515) 
19121 El Toro Rd ● Silverado, California 92676 ● (949) 459‐9600● com‐power.com
Shielded T‐series Coupling/Decoupling Networks (CDN) 
 
V
 
V
 
o
o
dBµV 
dBm 
120 
13 
130 
23 
140 
33 
 
 
Umr(150) 
Umr(150)
Volts 
dBµV 
0.167 
104.5 
0.5 
114 
1.67 
124.5 
Page 8 of 12
INSTRUCTION MANUAL
Umr(150) 
dBm 
‐2.5 
17.5 
REV021215