Baumer TA201 Operating Instructions Manual - Page 32

Browse online or download pdf Operating Instructions Manual for Measuring Instruments Baumer TA201. Baumer TA201 36 pages. Electronic

TA201
rpm measuring
Example
Example
Example
Differential
Extension / Compression
32
Imp./U.
1 rev. = 1 pulse
1 rev. = 1 pulse
1 rev. = 10 pulses
1 rev. = 10 pulses
Speed measurment
When using the TA 201 for speed measurement for m/min., the time
unit for F1 and F2 is set in lines 30 and 31 (rpm), and the number of
pulses per metre (scale) for F1 is located in line 25, for F2 in line 26.
Formula for calculation of the scale factor:
Circumf.
pulse/rev.
(Measur.
(encoder)
wheel)
0,5 m
0,5 m
5.2

Measurement functions

The measurement functions are described below. These functions
are selected in line 21. For monitoring two speeds or rpm values or
of one speed and one rpm value which you wish to display in the
form of ratio, differential or extension / compression value (discre-
pancy in percentage terms). In the event of a percentage display the
scale is set with the factor 100.
Formula
Rpm
Ratic
F1
F1=100 rpm
F2=200 rpm
F2
F1-F2
F1=200 rpm
F2=100 rpm
(F2-F1)
F1=100 rpm
F2=200 rpm
F1
F1=200 rpm
F2=100 rpm
www.baumerivo.com
Scale
(line 25, 26)
1
0.1
10
1
Scale
(line 25, 26)
1
1
= 2
0,5
50
50
= 100 none
0,5
Scale
(line 27)
10
1
100
1000
Decimal point
Display
(line 36, 37)
example
none
9999 (rpm)
0.0
999.9 (rpm)
none
9999 (rpm)
0.0
999.9 (rpm)
pulses/rev.
circumference
Decimal point
Display
(line 36, 37)
example
none
9999 (rpm)
9999 (rpm)
Decimal point
(line 38)
0.0
none
none
0.0
Display
0.5
100
100 (%)
-50 (%)
-50 (%)