BOQU SD Series Operating Instructions Manual - Page 11

Browse online or download pdf Operating Instructions Manual for Measuring Instruments BOQU SD Series. BOQU SD Series 15 pages. Online colormeter

Adjust and set the system time displayed on the current main
view,Year, month, day, hour, minute and second can be set。
7.7. Instrument calibration
When the deviation of the detected value is large, the instrument
can be corrected
Zero adjustment
1. In circulation,Connect clean water for zero adjustment
2. Enter the instrument calibration screen , Press on the
operation panel
adjustment",Press the "OK" key in the middle of the operation
panel to adjust zero,If the operation is unsuccessful, repeat the
operation。
Sample adjustment
Close the water inlet valve before sample adjustment and
correction,Pull out the water inlet quick connect hose,Short the
Operating instructions for SD Series Online Color Meter
, When the cursor flashes at "zero
10