water inlet and outlet with quick connect hose。As shown in the
figure below:
1. Open the cleaning port , Pour in the full-scale standard
sample solution provided by the manufacturer when leaving the
factory,Close the cleaning cover。
2. Enter the instrument calibration screen , Press on the
operation panel
adjustment",Press the "OK" key in the middle of the operation
panel to sample adjust ,If the operation is unsuccessful, repeat
the operation。
During zero adjustment or sample adjustment , It is
recommended to pour clean water or sample solution for many
times , Reduce the influence of residual liquid on correction
results。
After the above operations are completed, connect the
inlet and outlet pipes according to the normal installation and
Range:Maximum range setting。
Set 0:
Zero adjustment,After pouring in water, Tighten the
Operating instructions for SD Series Online Color Meter
, When the cursor flashes at "sample