Setup REC interval: The user can setup the REC interval by
pressing
. The interval setup will blink in LCD and it can be
REC
INTV
/
changed by
and
keys. Pressing these keys one time will
HOLD
ZERO
change the setup by one second and pressing these keys for more
than two seconds the value will change continuously. Once the
setup is done, the user needs to press the timer again to store the
value into the memory.
Start/Stop REC measuring DATA: When the function selector is
under voltage and low resistance ranges, the user can start the
REC function by pressing
key. The REC symbol will display in
REC
the LCD. The REC down counter will also be shown in LCD to
forecast the next data sampling time. When the second that the
recording occurs, a disk symbol will be shown to indicate this
process. When the function selector is under insulation ranges, the
user can set the REC mode only before the measuring start. If the
REC mode is set, a measuring will start the REC process
automatically.
Data Capture Function: When the user want to store a reading
on the LCD, he can press
key to enter data capture window.
The tester will automatically searched the next available capture
memory to store the reading or the user can browse up and down
by press the
and
keys. Once the address is chosen, the
/
HOLD
ZERO
user need to press
key to store reading. The user can exit
STORE
REC
data capture window by press
EXIT
key again.
Erase Data Memory: The user can erase all the memory in the
following process. Turn the tester off. Press and hold
key then
REC
turn on the tester.
will be shown to indicate the erasing is
under process. When it is done, message
will be shown in
the LCD and the user can release the key.