5.7. Physical Specifications
5.7.1.
The DP5 specification table is identical to that found in the "User Manual for Amptek's DP5
Product Family". The physical and power specifications are listed below.
Physical
Dimensions
Extender
Weight
5.8. Power Specifications
Characteristic
Symbol
Supply Voltage
V
in
Supply Current
I
in
Inrush Current
I
INRUSH
Input Capacitance C
IN
The power dissipated by the X-123 depends most strongly on the detector temperature, which is set in
software. At full cooling, ΔT=70 C, the X-123 draws about 0.7 A at 5 V, or 3.5 W. If the detector is not cooled
X-123 User Manual
7 x 10 cm x 2.5 cm (2.7 x 3.9 x 1 in) Excludes extender
1.5" (3.8 cm) standard
Options: None, 3/8", 5", 9", and vacuum flanges
200 g (7 oz)
Min
Typical
4.0
5.0
0.70
0.50
0.40
0.35
2
50
Max
Units
Conditions
5.5
V
0.9
A
V
= 5.0 V, initial cooldown
IN
A
V
= 5.0 V, steady state full cooling
IN
A
V
= 5.0 V, ΔT=70°C
IN
A
V
= 5.0 V, ΔT=50°C
IN
A
V
= 5.0 V, no cooling or bias
IN
A
<100
F
13