Table 201 – Guidance and manufacturer´s declaration – electromagnetic emission – for all EQUIPMENT AND SYSTEMS
(see 6.8.3.201 a) 3))
Guidance and manufacturer´s declaration – electromagnetic emission
1
The MACH LED 115 is intended for use in the electromagnetic environment specified below. The
2
customer or the user of the MACH LED 115 should assure that it is used in such an environment.
3
Emissions test
Harmonic emissions
7
IEC 61000-3-2
Voltage fluctuations /
8
flicker emissions
IEC 61000-3-3
RF emissions
12
CISPR 15
Table 202 – Guidance and manufacturer's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS (see
6.8.3.201 a) 6))
Guidance and manufacturer´s declaration – electromagnetic immunity
The MACH LED 115 is intended for use in the electromagnetic environment specified below. The customer
or the user of the MACH LED 115 should assure that it is used in such an environment.
Immunity test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
± 2 kV for power
Electrical fast
supply lines
transient / burst
IEC 61000-4-4
± 1 kV for input/output lines
± 1 kV differential
Surge
mode
IEC 61000-4-5
± 2 kV common mode
< 5 % U
(>95 % dip in U
for 0,5 cycle
40 % U
Voltage dips, short
(60 % dip in U
interruptions and
for 5 cycles
voltage variations
on power supply
70 % U
input lines
(30 % dip in U
IEC 61000-4-11
for 25 cycles
< 5 % U
(>95 % dip in U
for 5 sec
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
NOTE
U
is the a. c. mains voltage prior to application of the test level.
T
59060001

9. Electromagnetic compatibility

Compliance
Class C
Complies
Complies
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
not applicable
± 1 kV differential
mode
± 2 kV common mode
< 5 % U
T
)
(>95 % dip in U
T
for 0,5 cycle
40 % U
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
)
(30 % dip in U
T
for 25 cycles
< 5 % U
T
)
(>95 % dip in U
T
for 5 sec
30 A/m
Electromagnetic environment - guidance
The MACH LED 115 is suitable for use in all
establishments, including domestic establishments and those directly con-
nected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
The MACH LED 115 is not suitable for interconnection
with other equipment.
Compliance level
T
)
T
T
)
T
T
)
T
T
)
T
Edition 01
Dr. Mach
Lamps and Engineering
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital environment. If
the user of the
MACH LED 115 requires
continued operation during power
mains interruptions, it is recommended
that the MACH LED 115 be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
02.08.2012 / Bak Page 13/15