Dr. Mach LED 115 Instructions For Use Manual - Page 14

Browse online or download pdf Instructions For Use Manual for Medical Equipment Dr. Mach LED 115. Dr. Mach LED 115 15 pages. Spot light with led technology

Mach LED 115
Table 204 – Guidance and manufacturer´s declaration – electromagnetic immunity –
for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING (see 6.8.3.201 b))
The MACH LED 115 is intended for use in the electromagnetic environment specified below. The customer
or the user of the MACH LED 115 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 V
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from structures, objects
and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic envi-
ronment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in
which the MACH LED 115 is used exceeds the applicable RF compliance level above, the MACH LED 115 should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the MACH
LED 115.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
59060001
Guidance and manufacturer´s declaration – electromagnetic immunity
Compliance level
3 V
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the MACH LED 115 includ-
ing cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d
, 1
17
d
, 1
17
d
, 2
34
where p is the maximum output power rating of the trans-
mitter in watts (W) according to the transmitter manufactur-
er and d is the recommended separation distance in metres
b
(m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey,
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with the following symbol:
Edition 01
Dr. Mach
Lamps and Engineering
P
P
80 MHz to 800 MHz
P
800 MHz to 2,5 GHz
a
should be less than the
b
02.08.2012 / Bak Page 14/15