Andatech PRODIGY User Manual - Page 15
Browse online or download pdf User Manual for Test Equipment Andatech PRODIGY. Andatech PRODIGY 20 pages.
BLUETOOTH
Check the bluetooth printer status. If the printer is switched
on and connected, the State will display "Working". If the
State displays "None", ensure the printer is on and press
"MATCH" .
Note: It may take several seconds for the connection to
complete.
TROUBLESHOOTING
'CALIBRATION NEEDED' WARNING
When the unit starts up and shows the "Calibration Needed"
warning and alarm, it means the unit has passed its calibration
due date. The unit can continue to be used by touching the
screen, however it is strongly recommended that it be sent to
us for calibration.
BATTERY LIFE TOO SHORT
Use only alkaline batteries, and if using rechargeable batteries,
use either NiMH or rechargeable alkaline batteries.
TOUCH SCREEN NOT RESPONDING
If the touchscreen is not sensitive enough, calibrate the screen
by pressing and holding the On/Off button (right button)
immediately followed by the Function button (left button). Then
follow the on-screen instructions.
CLEAN AIR TEST CONTINUES TO DETECT
ALCOHOL
Let the device sit for 5 minutes to clear out any possible
residue. If the problem persists, contact Andatech.
15