Alert J5 Instruction Manual - Page 18
Browse online or download pdf Instruction Manual for Test Equipment Alert J5. Alert J5 20 pages.
MAINTENANCE & TROUBLESHOOTING
Troubleshooting Table
IF THE ALERT J5
Does not power on
Displays Not Activated
Cannot print serially (using
communication cable)
Is displaying irregular
results
Has a clogged or dirty
mouthpiece
Is using a printer that is
out of paper
Is displaying
PEAK ERROR
Requires service
p. 12/13
THEN
Change the batteries
Press the left button
• Check that the printer is on
• Check the cable connection
• Check the cable
• Change the mouthpiece. Use
part number 95-000130, to
order more mouthpieces from
ACS
• Eating, drinking or smoking 20
minutes prior to taking the test
can contaminate results. Wait
for at least 20 minutes and
repeat the test
• Check if a calibration is due
• Change the mouthpiece
• Use part number 95-000130
to order more mouthpieces
from ACS
• Replace the paper roll
• Use part number 95-000171
to order more rolls from ACS
• You are trying to calibrate the
device with a solution less
than 100 BAC
• Do not attempt to calibrate
yourself
Contact ACS for return procedure
ALERT
J5
TM