Drucker Diagnostics 614B Operator's Manual - Page 7
Browse online or download pdf Operator's Manual for Laboratory Equipment Drucker Diagnostics 614B. Drucker Diagnostics 614B 16 pages.
G
S
ENERAL
PECIFICATIONS
Nominal Speed (125 mm holders):
Nominal RCF (125 mm holders):
Nominal Speed (75/100 mm
holders):
Nominal RCF (75/100 mm
holders):
Maximum capacity:
Overall Dimensions (H x W x D):
Centrifuge Motor:
Nominal Acceleration Time:
Protection Breaker:
Timer:
Any use other than those specified by the Manufacturer is explicitly prohibited.
Maximum sample density is 1.15 grams / mL, (water density = 1.0 grams / mL)
S
L
ETUP
OCATION
1. Unpack the centrifuge and verify that all of the supplied
equipment is present.
2. Choose a setup location which meets the following criteria:
a.
A clearance height of 20" (50.8cm) is required to open the
lid.
b. The clearance envelope is the space around the centrifuge
which is required for safety. Choose a setup location which
will allow for a clearance envelope of at least 24" x 24"
(60.96cm x 60.96cm), (with the centrifuge at the center). No
person or hazardous material shall be permitted in the
clearance envelope during operation. The operator time
within the envelope shall be limited to the time necessary
for loading, unloading and centrifuge operation only.
c.
Proper ventilation is necessary to prevent the overheating of
samples as well as premature failure of the centrifuge.
Choose an area which will allow unencumbered air flow.
d. The centrifuge is designed to secure to the operating surface
by four suction feet. No adjustment is necessary for leveling
the centrifuge, however, the surface should be flat and level.
e.
Be sure the outlet is always within reach as the line cord is
the means of emergency disconnection!
Page | 3 Drucker Diagnostics +1-814-692-7661 - [email protected]
3,150 (± 100) RPM
1,200 (± 60) xg
3,250 (± 100) RPM
1,080 (± 60) xg
90 mL (6 x 15 mL)
8.75 in. x 11.75 in. x 14 in.
1/30 HP, AC
45 seconds
4 Amp. re–settable
Mechanical, 1 to 30 minutes