Operation

LEDs showing
exposure times

Activating the Handpiece

If the Power Level LED is off, briefly press the START
button to activate the hand piece. The Power Level LED
will come on and a short audible signal is emitted. The
hand piece is now ready for operation; the handpiece
displays the latest set exposure time setting.

Selection of Exposure Time

Press the TIME button to cycle through the following
settings from left to right. Press the TIME button again
to go back to 3s settings from the Extended Mode.
3s
6s
9s
Caution! High intensity light always involves the
production of heat. When utilizing the light in
Extended Mode (15s), do not keep the light in
one position. Longer exposure in the region of the pulp
can lead to injury or irritation of the pulp, which is why
the specified exposure times must not be exceeded.
Activating and Deactivating the
Light Emission
Briefly press the START button to activate the light
emission for the Exposure Time set as described above.
•  Once the emission has begun, the active Exposure
Time LEDs will count down every 3s with an acoustical
signal indicating the elapsed time, except for the
Extended Mode. When the emission is completed, the
Exposure Time LEDs will return as set before.
•  Example with 12s Exposure Time setting:
Beginning
3 sec
of Emission
Elapsed
ON Signal
1 beep
7
English
START button
Power Level LED
TIME button
Extended
12s
Mode (15s)
6 sec
9 sec
End of
Elapsed
Elapsed
Emission
2 beeps
3 beeps OFF Signal
•  For Extended Mode, an acoustical signal is emitted 
every 3 seconds for a total of 15 seconds.
•  If desired, press the START button to stop the light 
emission before the set time is over.

Tack Cure Function

Hold down the START button to activate the Tack Cure
function regardless of the set Exposure Time. The unit
emits a single one-second light pulse which allows for
a short exposure to tack cure brackets. The light will
return to standard operation when the START button is
no longer depressed.

Recommended Cure Techniques with

3M Unitek Adhesives

•     Rotate the light guide into the desired position 
for polymerization.
•       F or most 3M Unitek adhesives, reference the table 
below for recommended cure technique.
Appliance
Adhesive
Metal
Transbond
Brackets
Transbond
Ceramic
Change Adhesive,
Brackets
APC
Coated Brackets,
Buccal Tubes
APC
PLUS Adhesive
(direct bond)
Coated Brackets
Lingual
Transbond
Retainers
Transbond
Band Adhesive
Molar
Bands
Unitek
Multi-Cure Glass
Ionomer Band Cement
Indirect Trays
(Metal Brackets)
Transbond
LV Adhesive
Indirect Trays
(Ceramic
Brackets)
•  For other 3M Unitek adhesives not specifically 
mentioned, please refer to the adhesive Instructions
for Use.
•    I f using other light cure adhesive systems, consult 
the adhesive manufacturer's instructions.
•    T he light guide tip should be placed approximately 
1-2 mm above the appliance and pointed
perpendicular towards the tooth surface.
Ortholux
Luminous
Curing Light
3 seconds mesial x
XT Adhesive,
3 seconds distal
PLUS Color
3 seconds through
the bracket
II Adhesive
6 seconds mesial x
6 seconds occlusal
3 seconds mesial x
LR Adhesive
3 seconds distal
Plus
12 seconds
(3 seconds
per cusp)
6 seconds mesial
x 6 seconds distal
through the tray
Supreme
6 seconds through
the bracket
and tray