3M Unitek Ortholux Luminous Curing Ligh Instructions For Use Manual - Page 5

Browse online or download pdf Instructions For Use Manual for Medical Equipment 3M Unitek Ortholux Luminous Curing Ligh. 3M Unitek Ortholux Luminous Curing Ligh 16 pages. Fast curing cordless led light

3M Unitek Ortholux Luminous Curing Ligh Instructions For Use Manual
Classification: Protection class II,

Handpiece

Power supply:
Utilizable wavelength
range:
Wavelength peak:
Light intensity (between
400 and 515 nm):
Light emission area:
Intermittent
operation:
Total exposure time
with new, fully
charged battery:
Dimensions:
Weight:

Charger and Handpiece

Time to charge
empty battery:
Operating temperature: 10°C to 40°C (59°F to 104°F)
Relative humidity:
Atmospheric pressure: 700 hPa to 1060 hPa
Total height with
handpiece inserted in
the charger:

Transport and Storage Conditions:

Ambient temperature
range:
Relative humidity:
Atmospheric
pressure:
Subject to technical modification without prior notice.
5
English
lithium-ion battery, nominal
voltage 3.7 V
430-480 nm
455 nm +/- 10 nm
1600 mW/cm
(independent
2
of battery power level)
typically 42 mm
2
(optically active)
1 minute on, 15 minutes off
(ambient temperature of 40°C)
typically 7 minutes operating time
at ambient temperature (23°C)
typically 60 minutes
Diameter: 28 mm
Length: 270 mm
250 g (including light guide)
approximately 2.0 hours
30% to 75%
180 mm
-20°C to +40°C /
-4°F to +104°F
30% to 75%
700 hPa to 1060 hPa

Installation of the Unit

Factory Settings

The factory settings of the unit are as follows:
•  When first turned on the unit is in Extended Mode,
i.e. only the Power level LED is lit on the hand piece.

Initial Steps

Charger

•  Please ensure first that the voltage stated on the 
rating plate corresponds to the existing main supply
voltage. The rating plate is attached to the bottom
of the charger.
•  Place the charger on a level surface.
•  Connect the power cable of the charger to the 
power supply.
- The green LED on the charger showing that the unit
is ready for operation comes on. This shows that
the unit is ready for operation; please refer to the
section "Operating Status Display of the Charger."

Light Guide/Handpiece

•  Never place the handpiece in the charger without 
the battery inserted in the handpiece!
•  Place the eye shield on the front of the unit.
•  Autoclave the light guide prior to first use.
•  Insert the light guide firmly into the handpiece.
Removing and Inserting the Light Guide
from/into the Handpiece
•  The light guide has a magnetic holder. Remove the 
light guide from the handpiece by pulling it towards
the front.
•  Attach the light guide to the handpiece until it is 
firmly in place.