Classification: Protection class II,

Handpiece

Power supply:
Utilizable wavelength
range:
Wavelength peak:
Light intensity (between
400 and 515 nm):
Light emission area:
Intermittent
operation:
Total exposure time
with new, fully
charged battery:
Dimensions:
Weight:

Charger and Handpiece

Time to charge
empty battery:
Operating temperature: 10°C to 40°C (59°F to 104°F)
Relative humidity:
Atmospheric pressure: 700 hPa to 1060 hPa
Total height with
handpiece inserted in
the charger:

Transport and Storage Conditions:

Ambient temperature
range:
Relative humidity:
Atmospheric
pressure:
Subject to technical modification without prior notice.
5
English
lithium-ion battery, nominal
voltage 3.7 V
430-480 nm
455 nm +/- 10 nm
1600 mW/cm
(independent
2
of battery power level)
typically 42 mm
2
(optically active)
1 minute on, 15 minutes off
(ambient temperature of 40°C)
typically 7 minutes operating time
at ambient temperature (23°C)
typically 60 minutes
Diameter: 28 mm
Length: 270 mm
250 g (including light guide)
approximately 2.0 hours
30% to 75%
180 mm
-20°C to +40°C /
-4°F to +104°F
30% to 75%
700 hPa to 1060 hPa

Installation of the Unit

Factory Settings

The factory settings of the unit are as follows:
•  When first turned on the unit is in Extended Mode,
i.e. only the Power level LED is lit on the hand piece.

Initial Steps

Charger

•  Please ensure first that the voltage stated on the 
rating plate corresponds to the existing main supply
voltage. The rating plate is attached to the bottom
of the charger.
•  Place the charger on a level surface.
•  Connect the power cable of the charger to the 
power supply.
- The green LED on the charger showing that the unit
is ready for operation comes on. This shows that
the unit is ready for operation; please refer to the
section "Operating Status Display of the Charger."

Light Guide/Handpiece

•  Never place the handpiece in the charger without 
the battery inserted in the handpiece!
•  Place the eye shield on the front of the unit.
•  Autoclave the light guide prior to first use.
•  Insert the light guide firmly into the handpiece.
Removing and Inserting the Light Guide
from/into the Handpiece
•  The light guide has a magnetic holder. Remove the 
light guide from the handpiece by pulling it towards
the front.
•  Attach the light guide to the handpiece until it is 
firmly in place.