Apogee Instruments SQ-615 Owner's Manual - Page 8

Browse online or download pdf Owner's Manual for Accessories Apogee Instruments SQ-615. Apogee Instruments SQ-615 18 pages. Epar sensor
Also for Apogee Instruments SQ-615: Owner's Manual (18 pages)

Apogee Instruments SQ-615 Owner's Manual
Spectral Response
Mean spectral response measurements of
four replicate Apogee SQ-600 series ePAR
Sensors. Incremental spectral response
measurements were made at 10 nm
increments across a wavelength range of
370 to 800 nm in a monochromator with an
attached electric light source. Measured
spectral data from each quantum sensor
were refined and normalized by comparing
measured spectral response of the
monochromator/electric light combination
to measured spectral differences from a
quantum sensor reference.