Apogee Instruments SQ-615 Руководство пользователя - Страница 8

Просмотреть онлайн или скачать pdf Руководство пользователя для Аксессуары Apogee Instruments SQ-615. Apogee Instruments SQ-615 18 страниц. Epar sensor
Также для Apogee Instruments SQ-615: Руководство пользователя (18 страниц)

Apogee Instruments SQ-615 Руководство пользователя
Spectral Response
Mean spectral response measurements of
four replicate Apogee SQ-600 series ePAR
Sensors. Incremental spectral response
measurements were made at 10 nm
increments across a wavelength range of
370 to 800 nm in a monochromator with an
attached electric light source. Measured
spectral data from each quantum sensor
were refined and normalized by comparing
measured spectral response of the
monochromator/electric light combination
to measured spectral differences from a
quantum sensor reference.